Tyenaza ETCR2100A+

Tyenaza ETCR2100A+ Digital Clamp Resistance Tester User Manual

Model: ETCR2100A+

1. ആമുഖം

This manual provides essential instructions for the safe and effective operation of the Tyenaza ETCR2100A+ Digital Clamp Resistance Tester. This device is designed for accurate measurement of grounding resistance and circuit resistance, featuring a dustproof and shockproof structure for reliable performance in various industrial and departmental settings.

Please read this manual thoroughly before using the instrument to ensure proper handling and to prevent damage or injury.

Tyenaza ETCR2100A+ Digital Clamp റെസിസ്റ്റൻസ് ടെസ്റ്റർ

Image 1.1: The Tyenaza ETCR2100A+ Digital Clamp Resistance Tester, showing its overall design with the clamp jaw and control panel.

2 സുരക്ഷാ വിവരങ്ങൾ

Always observe the following safety precautions to avoid electric shock, injury, or damage to the instrument:

  • Do not use the instrument if it appears damaged or is operating abnormally.
  • പ്രവർത്തിക്കുന്നതിന് മുമ്പ് ബാറ്ററി കമ്പാർട്ട്മെന്റ് സുരക്ഷിതമായി അടച്ചിട്ടുണ്ടെന്ന് ഉറപ്പാക്കുക.
  • Do not attempt to measure live circuits beyond the specified voltage limits (CAT III 150V).
  • Avoid using the instrument in wet environments or during thunderstorms.
  • Refer to qualified personnel for any repairs or servicing.
  • Keep the instrument away from strong magnetic fields and electric fields exceeding specified limits.

3. പാക്കേജ് ഉള്ളടക്കം

പാക്കേജിൽ എല്ലാ ഇനങ്ങളും ഉണ്ടെന്ന് ഉറപ്പാക്കുക:

  • 1 x ETCR2100A+ Digital Clamp റെസിസ്റ്റൻസ് ടെസ്റ്റർ
  • 1 x Accessory (likely a test ring or similar calibration aid)
  • 1 x ഉപയോക്തൃ മാനുവൽ (ഈ പ്രമാണം)
  • 1 x സ്റ്റോറേജ് ബോക്സ്

Note: 1.5V AA batteries (x4) are required and are not included in the package.

ETCR2100A+ Digital Clamp Resistance Tester with included accessory

Image 3.1: The ETCR2100A+ tester shown with a standard test ring, which is typically included as an accessory for calibration or verification.

4. ഉൽപ്പന്ന സവിശേഷതകൾ

The ETCR2100A+ Digital Clamp Resistance Tester offers the following key features:

  • മൾട്ടിഫങ്ഷണൽ ഡിസ്പ്ലേ: Resistance and current values are displayed simultaneously on a new black luxury screen.
  • ഉയർന്ന കൃത്യത: Microprocessor-controlled for precise grounding resistance detection, utilizing fast filtering technology to minimize interference.
  • ഡാറ്റ മാനേജ്മെൻ്റ്: Features real-time clock, data storage for up to 99 sets, and data retrieval capabilities.
  • കേൾക്കാവുന്നതും ദൃശ്യപരവുമായ അലാറം: "Beep beep beep" alarm sound with adjustable threshold settings for resistance (1-199 Ω) and current (1-499mA).
  • യാന്ത്രിക പ്രവർത്തനങ്ങൾ: Automatic shutdown and fully automatic gear shifting for ease of use.
  • ശക്തമായ ഡിസൈൻ: Shockproof and dustproof structure, suitable for demanding industrial environments.
  • വിശാലമായ ആപ്ലിക്കേഷൻ: Ideal for telecommunications, power, meteorology, computer rooms, oil fields, power distribution lines, tower transmission lines, gas stations, factory grounding grids, and lightning rod systems.
ക്ലോസ് അപ്പ് view cl ന്റെamp jaws of the ETCR2100A+ tester

ചിത്രം 4.1: ഒരു ക്ലോസപ്പ് view of the robust clamp jaws, highlighting the double insulation and structural features designed for durability and safety.

5. സജ്ജീകരണം

5.1 ബാറ്ററി ഇൻസ്റ്റാളേഷൻ

The ETCR2100A+ requires four (4) 1.5V AA batteries (not included).

  1. ഉപകരണത്തിന്റെ പിൻഭാഗത്ത് ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ കണ്ടെത്തുക.
  2. Use a suitable tool (if necessary) to open the battery compartment.
  3. Insert four 1.5V AA batteries, ensuring correct polarity (+/-) as indicated inside the compartment.
  4. ബാറ്ററി കമ്പാർട്ട്മെൻ്റ് കവർ സുരക്ഷിതമായി അടയ്ക്കുക.

5.2 പ്രാരംഭ പവർ ഓൺ

ബാറ്ററി ഇൻസ്റ്റാൾ ചെയ്ത ശേഷം, അമർത്തുക ഓൺ/ഓഫ് button to power on the device. The LCD display will illuminate, indicating the device is ready for use.

6. പ്രവർത്തന നിർദ്ദേശങ്ങൾ

6.1 അടിസ്ഥാന അളവ്

To perform a grounding resistance test:

  1. ഉപകരണം ഓണാണെന്ന് ഉറപ്പാക്കുക.
  2. cl തുറക്കുകamp ട്രിഗർ അമർത്തി താടിയെല്ലുകൾ.
  3. cl സ്ഥാപിക്കുകamp around the grounding conductor or circuit to be measured. Ensure the jaws are fully closed around the conductor.
  4. The resistance value will be displayed on the LCD screen. The device features fully automatic shifting.
  5. The single measurement time is approximately 0.5 seconds.
ETCR2100A+ tester clamping an insulated ground wire

Image 6.1: The ETCR2100A+ tester shown clamping an insulated ground wire, demonstrating its capability to accurately test even insulated conductors.

6.2 Single Point Grounding System, Auxiliary Measurement

For single point grounding systems where direct clamp measurement of the lead wire is not feasible, auxiliary methods can be used. The two-point method is illustrated below:

Diagram illustrating the two-point auxiliary measurement method for single point grounding systems

Image 6.2: Diagram showing the two-point method for auxiliary grounding resistance measurement. It depicts a grounding rod (RA) connected to a single point, with a wire extending to a natural grounding point (e.g., water pipe, buried metal) for measurement.

ഈ രീതി ഉപയോഗിക്കുന്നതിന്:

  • Identify a natural grounding point with a low resistance value near the grounding point to be measured. Examples include a water pipe, fire hydrant, or other deeply buried metal structure.
  • Connect a wire from the point to be measured to this auxiliary grounding point.
  • Clamp the ETCR2100A+ around the connecting wire to perform the measurement.
ETCR2100A+ Digital Clamp Resistance Tester in an industrial substation environment

Image 6.3: The ETCR2100A+ tester positioned in an industrial substation, illustrating a typical application environment for grounding resistance measurement.

6.3 ഡാറ്റ സംഭരണവും വീണ്ടെടുക്കലും

The instrument can store up to 99 sets of measurement data.

  • To store data: After a measurement, press the എം.ഇ.എം button to save the current reading.
  • To retrieve data: Press the എം.ഇ.എം button repeatedly to cycle through stored data.

6.4 അലാറം പ്രവർത്തനം

The ETCR2100A+ features an audible and visual alarm for out-of-range measurements.

  • To activate/deactivate the alarm: Press the AL ബട്ടൺ.
  • To set alarm thresholds: Refer to the device's on-screen menu or specific instructions for setting resistance (1-199 Ω) and current (1-499mA) alarm limits.
Hand holding the ETCR2100A+ Digital Clamp റെസിസ്റ്റൻസ് ടെസ്റ്റർ

Image 6.4: A hand holding the ETCR2100A+ tester, demonstrating its ergonomic design and ease of handling during operation.

7. പരിപാലനം

7.1 വൃത്തിയാക്കൽ

Wipe the instrument's exterior with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the device is powered off before cleaning.

7.2 സംഭരണം

When not in use for extended periods, remove the batteries to prevent leakage. Store the instrument in its provided storage box in a cool, dry place, away from direct sunlight and extreme temperatures.

7.3 കാലിബ്രേഷൻ

Regular calibration by qualified personnel is recommended to maintain measurement accuracy, especially after prolonged use or if accuracy is suspected to be compromised.

8. പ്രശ്‌നപരിഹാരം

പ്രശ്നംസാധ്യമായ കാരണംപരിഹാരം
ഉപകരണം പവർ ഓണാക്കുന്നില്ല.ഡെഡ് അല്ലെങ്കിൽ തെറ്റായി ഇൻസ്റ്റാൾ ചെയ്ത ബാറ്ററികൾ.ശരിയായ പോളാരിറ്റി ഉറപ്പാക്കിക്കൊണ്ട് ബാറ്ററികൾ പുതിയ 1.5V AA ബാറ്ററികൾ ഉപയോഗിച്ച് മാറ്റിസ്ഥാപിക്കുക.
കൃത്യമല്ലാത്ത വായനകൾ.Jaws not fully closed; interference from external fields; dirty jaws; need for calibration.Ensure jaws are fully closed. Move away from strong magnetic/electric fields. Clean jaws. Consider professional calibration.
അലാറം തുടർച്ചയായി മുഴങ്ങുന്നു.Measurement exceeds set alarm threshold.Check the measured value against the set alarm limits. Press the AL button to silence or deactivate the alarm.
ഡിസ്പ്ലേ ശൂന്യമാണ് അല്ലെങ്കിൽ മിന്നിമറയുന്നു.Low battery power; environmental conditions (extreme temperature/humidity).Replace batteries. Operate within specified temperature and humidity ranges (-20 ℃ to 55 ℃; 10% RH to 90% RH).

9 സ്പെസിഫിക്കേഷനുകൾ

മോഡൽETCR2100A+
ഫംഗ്ഷൻGrounding resistance test, circuit resistance test, leakage current test
പ്രതിരോധ ശ്രേണി0.01 Ω മുതൽ 200 Ω വരെ
റെസിസ്റ്റൻസ് റെസല്യൂഷൻ0.001 Ω
നിലവിലെ ശ്രേണിNone (for leakage current test, the range is 1-499mA for alarm setting)
ഡാറ്റ സംഭരണം99 സെറ്റുകൾ
Alarm Threshold Setting Range (Resistance)1-199 Ω
Alarm Threshold Setting Range (Current)1-499mA
പ്രവർത്തന താപനിലയും ഈർപ്പവും-20 ℃ to 55 ℃; 10% RH to 90% RH
എൽസിഡി ഡിസ്പ്ലേ4 digit LCD digital display
ഡിസ്പ്ലേ വലിപ്പംApprox. 47 x 28.5mm / 1.85 x 1.12in (L x W)
Jaw Size (Round)Approx. 32mm / 1.26in diameter
Jaw Opening Size (Long)Approx. 32mm / 1.26in
Clamp Size (Long Jaw)ഏകദേശം 260 x 90 x 56 മിമി / 10.24 x 3.54 x 2.2 ഇഞ്ച്
സംരക്ഷണ നിലഇരട്ട ഇൻസുലേഷൻ
ഘടനാപരമായ സവിശേഷതകൾClamp shaped CT
ഗിയർ ഷിഫ്റ്റിംഗ്Fully automatic shifting
ബാഹ്യ കാന്തിക മണ്ഡലം<40A/m
ബാഹ്യ ഇലക്ട്രിക് ഫീൽഡ്<1V/m
Single Measurement Time0.5 സെക്കൻഡ്
Resistance Measurement FrequencyGreater than 1KHz
Tested Current Frequency50/60Hz automatic
വൈദ്യുതി വിതരണം4 x 1.5V AA ബാറ്ററികൾ (ഉൾപ്പെടുത്തിയിട്ടില്ല)
മെറ്റീരിയൽഎബിഎസ്
ഇനത്തിൻ്റെ ഭാരം2.13 കിലോഗ്രാം / 4.69 പൗണ്ട്

10. വാറൻ്റിയും പിന്തുണയും

വാറന്റി വിവരങ്ങൾ, സാങ്കേതിക പിന്തുണ അല്ലെങ്കിൽ സേവന അന്വേഷണങ്ങൾ എന്നിവയ്‌ക്കായി, ദയവായി നിങ്ങളുടെ റീട്ടെയിലറെയോ നിർമ്മാതാവിനെയോ നേരിട്ട് ബന്ധപ്പെടുക. വാങ്ങിയതിന്റെ തെളിവായി നിങ്ങളുടെ വാങ്ങൽ രസീത് സൂക്ഷിക്കുക.

Manufacturer: Tyenaza

© 2023 Tyenaza. All rights reserved.