1. ആമുഖം
This manual provides detailed instructions for the proper operation, maintenance, and troubleshooting of the INSTRUKART Lutron TU-2016 Turbidity Meter. Please read this manual thoroughly before using the device to ensure accurate measurements and safe operation.
The Lutron TU-2016 is designed for measuring turbidity in various water sources, including lakes, water tanks, rivers, and mangroves. It features a wide measurement range, high resolution, and a user-friendly interface, adhering to ISO 7027 standards.
2 സുരക്ഷാ വിവരങ്ങൾ
- Always handle the meter and its components with care to prevent damage.
- Do not expose the device to extreme temperatures, humidity, or direct sunlight for prolonged periods.
- Ensure the battery compartment is properly sealed to prevent water ingress.
- Keep the exterior of the test bottle clean and dry before insertion into the meter.
- Use only specified standard solutions for calibration.
- Dispose of batteries and solutions according to local regulations.
3. പാക്കേജ് ഉള്ളടക്കം
അൺപാക്ക് ചെയ്യുമ്പോൾ, താഴെ ലിസ്റ്റ് ചെയ്തിരിക്കുന്ന എല്ലാ ഇനങ്ങളും നിലവിലുണ്ടെന്നും നല്ല നിലയിലാണെന്നും ഉറപ്പാക്കുക:
- 1 Unit of TU-2016 Turbidity Meter
- 0 NTU standard solution
- 10 NTU standard solution
- Empty testing bottle (sample cuvette)
- Hard Carrying case
- ഇൻസ്ട്രക്ഷൻ മാനുവൽ
- ഫാക്ടറി കാലിബ്രേഷൻ സർട്ടിഫിക്കറ്റ്

ചിത്രം 3.1: Contents of the TU-2016 Turbidity Meter package, neatly organized within its hard carrying case.

ചിത്രം 3.2: The TU-2016 Turbidity Meter along with its essential accessories for measurement and maintenance.
4. ഉൽപ്പന്നം കഴിഞ്ഞുview
The Lutron TU-2016 Turbidity Meter is a portable, microprocessor-controlled device designed for precise turbidity measurements. Key features include:
- ISO 7027 Compliance: Designed to meet international standards for turbidity measurement.
- അളക്കൽ യൂണിറ്റ്: NTU (Nephelometric Turbidity Unit).
- Wide Auto Measurement Range: 0 മുതൽ 1,000 NTU വരെ.
- ഉയർന്ന മിഴിവ്: 0.01 NTU (for 0.00 to 50.00 NTU range) and 1 NTU (for 50 to 1,000 NTU range).
- Unique Optics Structure: Enables accurate readings from low to high turbidity levels.
- ഉപയോക്തൃ-സൗഹൃദ ഇൻ്റർഫേസ്: Four operation buttons and three calibration points for easy use.
- Jumbo LCD Display: Large display for clear and easy readability.
- Microprocessor Circuit: Ensures maximum accuracy and provides special functions.
- ബാറ്ററി പ്രവർത്തിപ്പിക്കുന്നത്: Ideal for field and on-site testing.
- ഡാറ്റ ഹോൾഡ് പ്രവർത്തനം: Freezes the displayed value.
- മെമ്മറി റികോൾ: Records and recalls Maximum and Minimum readings.
- യാന്ത്രിക ഷട്ട്-ഓഫ്: ബാറ്ററി ലൈഫ് സംരക്ഷിക്കുന്നു.
- മോടിയുള്ള ഡിസൈൻ: Heavy-duty and compact housing with a hard carrying case.

ചിത്രം 4.1: ഫ്രണ്ട് view of the TU-2016 Turbidity Meter, highlighting the display and control panel.

ചിത്രം 4.2: The TU-2016 Turbidity Meter held in hand, illustrating its compact and portable design.
5. സജ്ജീകരണം
5.1 ബാറ്ററി ഇൻസ്റ്റാളേഷൻ
The meter operates on six DC 1.5 V AAA (UM4) batteries. To install or replace batteries:
- മീറ്ററിന്റെ പിൻഭാഗത്ത് ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ കണ്ടെത്തുക.
- Open the cover and insert six AAA batteries, observing the correct polarity (+/-).
- ബാറ്ററി കമ്പാർട്ട്മെൻ്റ് കവർ സുരക്ഷിതമായി അടയ്ക്കുക.
5.2 പ്രാരംഭ പവർ ഓൺ
അമർത്തുക പവർ button to turn on the meter. The LCD display will illuminate and show a brief self-test before entering measurement mode.
6. പ്രവർത്തന നിർദ്ദേശങ്ങൾ
6.1 Preparing a Sample
- Ensure the empty testing bottle (cuvette) is clean and dry. Wipe the exterior with a lint-free cloth.
- Fill the testing bottle with the water sample to be measured. Avoid air bubbles.
- Securely cap the bottle.
6.2 ഒരു അളവ് എടുക്കുന്നു
- എസ് തുറക്കുകample compartment lid on the top of the meter.
- Insert the prepared sample bottle into the compartment, ensuring it is seated correctly.
- എസ് അടയ്ക്കുകample കമ്പാർട്ട്മെൻ്റ് ലിഡ്.
- അമർത്തുക TEST/CAL button to initiate a measurement. The turbidity reading will appear on the LCD display.

ചിത്രം 6.1: Opening the sample കമ്പാർട്ട്മെൻ്റ് ലിഡ്.

ചിത്രം 6.2: Inserting a sample bottle into the meter.

ചിത്രം 6.3: The meter with the sample compartment open, ready for sample insertion.
6.3 ഡാറ്റ ഹോൾഡ് പ്രവർത്തനം
ഡിസ്പ്ലേയിൽ നിലവിലെ റീഡിംഗ് മരവിപ്പിക്കാൻ, പിടിക്കുക button. Press it again to release and resume live measurement.
6.4 Memory Recall (MAX/MIN)
The meter can record and recall the maximum and minimum turbidity readings:
- അമർത്തുക REC button once to display the maximum recorded value.
- അമർത്തുക REC button again to display the minimum recorded value.
- അമർത്തുക REC button a third time to exit memory recall mode.
6.5 ഓട്ടോ ഷട്ട്-ഓഫ്
The meter features an auto shut-off function to conserve battery life. If no button is pressed for a certain period, the meter will automatically power off. This can be manually overridden by pressing the പവർ ഉപകരണം ഓഫ് ചെയ്യാനുള്ള ബട്ടൺ.
7. കാലിബ്രേഷൻ
The TU-2016 features a three-point push-button calibration system. Regular calibration ensures the accuracy of your measurements. Use the provided 0 NTU and 10 NTU standard solutions for calibration.
- മീറ്റർ ഓണാക്കുക.
- Insert the 0 NTU standard solution into the sampലെ കമ്പാർട്ട്മെന്റ്.
- അമർത്തിപ്പിടിക്കുക TEST/CAL button until the calibration mode is indicated on the display. Follow the on-screen prompts to calibrate the 0 NTU point.
- Remove the 0 NTU solution and insert the 10 NTU standard solution.
- Repeat the calibration process for the 10 NTU point.
- The third calibration point (if applicable) can be performed similarly with a higher NTU standard solution if available and required for your specific application.
- After successful calibration, the meter will return to measurement mode.
Note: Refer to the Factory Calibration Certificate for details on the initial calibration performed by the manufacturer.
8. പരിപാലനം
8.1 മീറ്റർ വൃത്തിയാക്കൽ
- മീറ്ററിന്റെ പുറംഭാഗം മൃദുവായ, ഡി-ടൈപ്പ് ഉപയോഗിച്ച് തുടയ്ക്കുക.amp തുണി. ഉരച്ചിലുകളുള്ള ക്ലീനറുകളോ ലായകങ്ങളോ ഉപയോഗിക്കരുത്.
- എസ് സൂക്ഷിക്കുകample compartment clean and free of debris.
- Ensure the exterior of the test bottles are always clean and dry before use.
8.2 സംഭരണം
- When not in use, store the meter and its accessories in the provided hard carrying case.
- നേരിട്ടുള്ള സൂര്യപ്രകാശത്തിൽ നിന്നും തീവ്രമായ താപനിലയിൽ നിന്നും അകന്ന് തണുത്തതും വരണ്ടതുമായ സ്ഥലത്ത് സൂക്ഷിക്കുക.
- ദീർഘനേരം സൂക്ഷിക്കുകയാണെങ്കിൽ, ചോർച്ച തടയാൻ ബാറ്ററികൾ നീക്കം ചെയ്യുക.
9. പ്രശ്നപരിഹാരം
| പ്രശ്നം | സാധ്യമായ കാരണം | പരിഹാരം |
|---|---|---|
| മീറ്റർ പവർ ഓൺ ആകുന്നില്ല. | ഡെഡ് അല്ലെങ്കിൽ തെറ്റായി ഇൻസ്റ്റാൾ ചെയ്ത ബാറ്ററികൾ. | ബാറ്ററികൾ മാറ്റി സ്ഥാപിക്കുക, ശരിയായ പോളാരിറ്റി ഉറപ്പാക്കുക. |
| കൃത്യമല്ലാത്ത വായനകൾ. |
|
|
| Display shows "Err" or similar error. | Measurement range exceeded or internal error. | എസ് ഉറപ്പാക്കുകample is within 0-1000 NTU range. Power cycle the device. If error persists, contact support. |
10 സ്പെസിഫിക്കേഷനുകൾ
| പരാമീറ്റർ | വിശദാംശങ്ങൾ |
|---|---|
| സർക്യൂട്ട് | Custom one-chip of microprocessor LSI circuit |
| പ്രദർശിപ്പിക്കുക | LCD വലിപ്പം: 41 mm x 34 mm |
| അളക്കൽ ശ്രേണി | 0.00 to 50.00 NTU, 50 to 1,000 NTU (Auto range) |
| റെസലൂഷൻ | 0.01 NTU (0.00 to 50.00 NTU), 1 NTU (50 to 1,000 NTU) |
| കൃത്യത | ± 5 % F.S. or ± 0.5 NTU, whichever is greater |
| പ്രകാശ സ്രോതസ്സ് | എൽഇഡി, 850 എൻഎം |
| ഡിറ്റക്ടർ | ഫോട്ടോ ഡയോഡ് |
| സ്റ്റാൻഡേർഡ് പാലിക്കൽ | Meet ISO 7027 |
| പ്രതികരണ സമയം | 10 സെക്കൻഡിൽ കുറവ് |
| Sample Volume Needed | 10 മി.ലി |
| ഡാറ്റ ഹോൾഡ് | ഡിസ്പ്ലേ റീഡിംഗ് ഫ്രീസ് ചെയ്യുക |
| മെമ്മറി റികോൾ | പരമാവധി & കുറഞ്ഞ മൂല്യം |
| ഡിസ്പ്ലേ എസ്ampലിംഗ് സമയം | ഏകദേശം. 1 സെക്കൻഡ് |
| പവർ ഓഫ് | പുഷ് ബട്ടൺ ഉപയോഗിച്ച് സ്വയമേവ ഷട്ട് ഓഫ് ചെയ്യുക അല്ലെങ്കിൽ മാനുവൽ ഓഫ് ചെയ്യുക |
| പ്രവർത്തന താപനില | 0 മുതൽ 50 ഡിഗ്രി സെൽഷ്യസ് വരെ |
| പ്രവർത്തന ഹ്യുമിഡിറ്റി | 85% RH-ൽ കുറവ് |
| വൈദ്യുതി വിതരണം | DC 1.5 V battery (UM4, AAA) x 6 PCs |
| Power Current (Standby) | ഏകദേശം. DC 3.5 mA |
| Power Current (Testing) | ഏകദേശം. DC 36 mA |
| ഭാരം | 320 g Approx. (0.71 lbs) |
| അളവുകൾ | 155 x 76 x 62 മിമി (6.1 x 2.99 x 2.44 ഇഞ്ച്) |
11. പിന്തുണയും വാറണ്ടിയും
For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact INSTRUKART customer support. Refer to your purchase documentation for specific warranty terms and contact information.
The product comes with a Factory Calibration Certificate, ensuring its initial accuracy and compliance with international standards.