1. ആമുഖം
The HORIBA LAQUA EC210-K is a robust, handheld, waterproof meter designed for precise measurement of conductivity, total dissolved solids (TDS), salinity, and temperature in various field applications. This digital meter provides reliable performance with advanced features such as automatic and manual calibration, adjustable parameters, and extensive data logging capabilities. Its durable construction ensures long-lasting use in challenging environments.
2 സുരക്ഷാ വിവരങ്ങൾ
Please read this manual thoroughly before operating the LAQUA EC210-K meter to ensure safe and correct usage. Failure to follow instructions may result in damage to the instrument or personal injury.
- Always handle the meter and sensor with care.
- Do not immerse the meter in liquids beyond its specified IP67 rating.
- Keep calibration solutions and samples away from children and pets. Follow safety data sheets for chemical handling.
- Use only specified batteries (AA) and dispose of them properly.
- Do not attempt to disassemble or repair the meter yourself. Refer to qualified service personnel.
3. പാക്കേജ് ഉള്ളടക്കം
Verify that all items are present and undamaged upon unpacking. The standard package for the HORIBA LAQUA EC210-K includes:
- LAQUA EC210-K Handheld Meter
- 9383-10D Conductivity Sensor
- Calibration Solutions (various concentrations)
- 2 AA ബാറ്ററികൾ
- Built-in Electrode Stand
- Foldable Meter Stand
- Durable Carrying Case
- നിർദ്ദേശ മാനുവൽ (ഈ പ്രമാണം)

Image 3.1: Contents of the HORIBA LAQUA EC210-K kit, including the meter, sensor, calibration solutions, and accessories, neatly organized within its protective carrying case.
4. സജ്ജീകരണം
4.1. ബാറ്ററി ഇൻസ്റ്റാളേഷൻ
- മീറ്ററിന്റെ പിൻഭാഗത്ത് ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ കണ്ടെത്തുക.
- Open the cover by turning the latch or unscrewing as indicated.
- ശരിയായ പോളാരിറ്റി (+/-) ഉറപ്പാക്കിക്കൊണ്ട് രണ്ട് (2) AA ബാറ്ററികൾ ഇടുക.
- വാട്ടർപ്രൂഫ് സീൽ നിലനിർത്താൻ ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ സുരക്ഷിതമായി അടയ്ക്കുക.
4.2. സെൻസർ കണക്ഷൻ
- Align the connector of the 9383-10D conductivity sensor with the sensor input port on the top of the meter.
- Gently push and twist the connector until it is securely fastened. Do not overtighten.

Image 4.1: The HORIBA LAQUA EC210-K handheld meter with its 9383-10D conductivity sensor securely connected, ready for use.
5. പ്രവർത്തന നിർദ്ദേശങ്ങൾ
5.1. പവർ ഓൺ/ഓഫ്
- പവർ ഓൺ ചെയ്യാൻ: അമർത്തിപ്പിടിക്കുക പവർ button (⑂) until the display illuminates.
- പവർ ഓഫ് ചെയ്യാൻ: അമർത്തിപ്പിടിക്കുക പവർ button (⑂) again until the display turns off.
5.2. അടിസ്ഥാന അളവെടുപ്പ്
- Ensure the sensor is clean and properly connected.
- മീറ്ററിൽ പവർ.
- അമർത്തുക മോഡ് button to cycle through measurement parameters (Conductivity, TDS, Salinity, Temperature).
- Immerse the sensor in the sample solution, ensuring the sensor tip is fully submerged and free of air bubbles.
- Wait for the reading to stabilize. The meter features auto-stable and auto-hold functions for convenience.
- Record the measurement displayed on the custom LCD with backlight.
5.3. Display and Button Functions
The meter features a clear LCD with backlight for easy reading in various lighting conditions. Key buttons include:
- മോഡ്: Cycles through measurement parameters (Conductivity, TDS, Salinity, Temperature).
- CAL: Initiates calibration mode.
- MEAS: Returns to measurement mode.
- ഡാറ്റ: Accesses data logging and memory functions.
- സെറ്റ്: Enters setup menu for parameter adjustments.
- ഇഎൻടി: Confirms selections or enters values.
- മുകളിലേക്ക്/താഴേക്ക് അമ്പടയാളങ്ങൾ: Navigate menus and adjust values.

ചിത്രം 5.1: മുൻഭാഗം view of the HORIBA LAQUA EC210-K meter, highlighting its custom LCD display and the arrangement of control buttons for intuitive operation.
6. കാലിബ്രേഷൻ
Regular calibration is essential for accurate measurements. The LAQUA EC210-K supports both automatic and manual calibration procedures.
6.1. EC Calibration
- അമർത്തുക CAL കാലിബ്രേഷൻ മോഡിൽ പ്രവേശിക്കാൻ ബട്ടൺ.
- Follow the on-screen prompts to select automatic or manual calibration.
- For automatic calibration, immerse the sensor in the appropriate standard solution. The meter will automatically recognize the solution and calibrate.
- For manual calibration, immerse the sensor in a standard solution and use the arrow buttons to adjust the reading to match the known value. Press ഇഎൻടി സ്ഥിരീകരിക്കാൻ.
- The meter allows adjustment of reference temperature, temperature coefficient, and cell constant for precise measurements.
6.2. TDS and Salinity Calibration
- The meter provides EC to TDS conversion with four selectable TDS calibration curves.
- Two salinity calibration curves are available for accurate salinity measurements.
- റഫർ ചെയ്യുക സെറ്റ് menu for selecting and configuring these curves.
7. ഡാറ്റ മാനേജ്മെൻ്റ്
The LAQUA EC210-K includes features for efficient data handling:
- Auto Log Data: Automatically records measurement data at set intervals.
- ബിൽറ്റ്-ഇൻ മെമ്മറി: Stores up to 500 data points, allowing for comprehensive record-keeping.
- Auto Stable/Auto Hold: Ensures stable readings are captured and held on the display.
- Real-time Measurement: Provides continuous display of current readings.
സംഭരിച്ച ഡാറ്റ ആക്സസ് ചെയ്യാൻ, ഡാറ്റ button and navigate using the arrow keys.
8. പരിപാലനം
Proper maintenance extends the life and accuracy of your meter.
8.1. Cleaning the Meter and Sensor
- Meter: Wipe the meter body with a soft, damp തുണി. ഉരച്ചിലുകളുള്ള ക്ലീനറുകളോ ലായകങ്ങളോ ഉപയോഗിക്കരുത്.
- Sensor: Rinse the sensor thoroughly with distilled or deionized water after each use. For stubborn deposits, refer to the sensor-specific cleaning instructions provided with the 9383-10D sensor.
8.2. Sensor Storage
Store the sensor according to its specific instructions, typically in a storage solution or with a protective cap to prevent drying out.
8.3. ബാറ്ററി മാറ്റിസ്ഥാപിക്കൽ
ഡിസ്പ്ലേയിൽ ബാറ്ററി കുറവാണെന്ന് കാണിക്കുമ്പോൾ ബാറ്ററികൾ മാറ്റിസ്ഥാപിക്കുക. ബാറ്ററി ഇൻസ്റ്റാളേഷൻ നിർദ്ദേശങ്ങൾക്ക് സെക്ഷൻ 4.1 കാണുക.
8.4. ഈട്
The LAQUA EC210-K features an IP67 waterproof and dustproof rating, and a shock and scratch-resistant non-slip housing. While designed for rugged use, avoid unnecessary drops or impacts.
9. പ്രശ്നപരിഹാരം
If you encounter issues with your LAQUA EC210-K meter, consult the following table for common problems and solutions.
| പ്രശ്നം | സാധ്യമായ കാരണം | പരിഹാരം |
|---|---|---|
| മീറ്റർ ഓൺ ആകുന്നില്ല. | മരിച്ചതോ തെറ്റായി ഇൻസ്റ്റാൾ ചെയ്തതോ ആയ ബാറ്ററികൾ | ബാറ്ററികൾ മാറ്റി സ്ഥാപിക്കുക, ശരിയായ പോളാരിറ്റി ഉറപ്പാക്കുക. |
| കൃത്യമല്ലാത്ത വായനകൾ | Sensor dirty or damaged; Meter out of calibration | Clean the sensor; Perform calibration (Section 6). |
| No reading displayed | Sensor not connected properly; Sensor damaged | Check sensor connection; Replace sensor if damaged. |
| ഡിസ്പ്ലേ മങ്ങിയതോ മിന്നിമറയുന്നതോ ആണ് | കുറഞ്ഞ ബാറ്ററി പവർ | ബാറ്ററികൾ മാറ്റിസ്ഥാപിക്കുക. |
If the problem persists after attempting these solutions, please contact HORIBA customer support.
10 സ്പെസിഫിക്കേഷനുകൾ
| പരാമീറ്റർ | സ്പെസിഫിക്കേഷൻ |
|---|---|
| മോഡലിൻ്റെ പേര് | LAQUA EC210-K |
| അളക്കൽ പാരാമീറ്ററുകൾ | Conductivity, Resistivity, Salinity, Total Dissolved Solids (TDS), Temperature |
| വാട്ടർപ്രൂഫ്/ഡസ്റ്റ് പ്രൂഫ് റേറ്റിംഗ് | IP67 |
| പവർ ഉറവിടം | 2 AA ബാറ്ററികൾ |
| ബാറ്ററി ലൈഫ് | 500 മണിക്കൂറിലധികം |
| ഡിസ്പ്ലേ തരം | Custom LCD with Backlight |
| മെമ്മറി | 500 ഡാറ്റ പോയിൻ്റുകൾ വരെ |
| ബാഹ്യ മെറ്റീരിയൽ | പോളികാർബണേറ്റ് (PC) |
| ഉൾപ്പെടുത്തിയ ഘടകങ്ങൾ | Meter and Sensor (9383-10D) |
| ഉയർന്ന താപനില റേറ്റിംഗ് | 130 ഡിഗ്രി സെൽഷ്യസ് |
| ഇനത്തിൻ്റെ ദൈർഘ്യം | 7 ഇഞ്ച് |
11. വാറൻ്റിയും പിന്തുണയും
11.1. വാറൻ്റി വിവരങ്ങൾ
The HORIBA LAQUA EC210-K meter comes with a 3 വർഷത്തെ വാറൻ്റി from the date of purchase, covering defects in materials and workmanship. This warranty does not cover damage caused by misuse, accident, unauthorized repair, or normal wear and tear. Please retain your proof of purchase for warranty claims.
11.2. ഉപഭോക്തൃ പിന്തുണ
For technical assistance, troubleshooting, or warranty service, please contact HORIBA Instruments Incorporated customer support. You can find more information and contact details by visiting the official HORIBA store or webസൈറ്റ്.
സന്ദർശിക്കുക HORIBA Store on Amazon.