HORIBA LAQUA EC220-K ഹാൻഡ്‌ഹെൽഡ് EC മീറ്റർ ഇൻസ്ട്രക്ഷൻ മാനുവൽ

മോഡൽ: LAQUA EC220-K

1. ആമുഖം

The HORIBA LAQUA EC220-K is a robust, handheld, and waterproof meter designed for precise water quality testing in various environments. This device accurately measures conductivity (EC), resistivity (RES), salinity (SAL), total dissolved solids (TDS), and temperature. Its durable construction and user-friendly features make it suitable for both laboratory and outdoor field applications.

പ്രധാന സവിശേഷതകളിൽ ഇവ ഉൾപ്പെടുന്നു:

  • വാട്ടർപ്രൂഫ് ഡിസൈൻ: IP67 rated for dust and water resistance, ensuring reliability in challenging conditions.
  • വിപുലമായ പ്രവർത്തനം: Offers automatic and manual EC calibration, adjustable reference temperature, temperature coefficient, cell constant, and selectable measurement units.
  • Expansive Measurement Ranges: Capable of measuring conductivity, resistivity, salinity, total dissolved solids, and temperature with high resolution and accuracy.
  • നീണ്ടുനിൽക്കുന്ന നിർമ്മാണം: Features a shock and scratch-resistant non-slip casing, a built-in electrode stand, and a foldable meter stand.
  • User-friendly Features: Includes auto log data, built-in memory for up to 500 data points, auto stable, auto hold, real-time measurement modes, and a custom LCD with backlight for clear readings.

2 സുരക്ഷാ വിവരങ്ങൾ

Please read and understand all safety precautions before operating the HORIBA LAQUA EC220-K meter. Failure to do so may result in injury or damage to the instrument.

  • Do not attempt to disassemble or modify the meter. This may void the warranty and cause damage.
  • Ensure the battery compartment is securely sealed to maintain the IP67 rating, especially when operating in wet conditions.
  • Handle the sensor with care. Avoid dropping or subjecting it to excessive force.
  • Keep the meter and sensor away from strong magnetic fields or corrosive chemicals unless specifically designed for such exposure.
  • Use only specified batteries (2 AA) for operation.

3. പാക്കേജ് ഉള്ളടക്കം

പായ്ക്ക് അൺപാക്ക് ചെയ്യുമ്പോൾ എല്ലാ ഇനങ്ങളും ഉണ്ടെന്ന് ഉറപ്പാക്കുക:

  • HORIBA LAQUA EC220-K Meter
  • EC Sensor (Model 9383-10D)
  • Calibration Solutions (as packaged)
  • 2 AA ബാറ്ററികൾ
  • ഇൻസ്ട്രക്ഷൻ മാനുവൽ
  • ചുമക്കുന്ന കേസ്

4. ഉൽപ്പന്നം കഴിഞ്ഞുview

HORIBA LAQUA EC220-K Handheld EC Meter with attached sensor

The image displays the HORIBA LAQUA EC220-K handheld meter in black, with its digital LCD showing 'COND', '25.0°C ATC', and a reading of '1413 µS/cm'. The meter features control buttons for CAL, MEAS, DATA, MODE, SET, ENT, and power. An EC sensor, labeled HORIBA, is connected to the right side of the meter.

4.1. Meter Components

  • LCD ഡിസ്പ്ലേ: Shows measurement readings, temperature, battery status, and operational icons.
  • CAL ബട്ടൺ: കാലിബ്രേഷൻ മോഡ് ആരംഭിക്കുന്നു.
  • MEAS ബട്ടൺ: Switches to measurement mode.
  • DATA Button: Accesses stored data.
  • മോഡ് ബട്ടൺ: Cycles through measurement parameters (EC, TDS, SAL, RES).
  • SET ബട്ടൺ: Enters setup menu for advanced settings.
  • ENT Button: Confirms selections or enters values.
  • മുകളിലേക്കും താഴേക്കുമുള്ള അമ്പടയാളങ്ങൾ: Navigate menus and adjust values.
  • പവർ ബട്ടൺ: Turns the meter on or off.
  • സെൻസർ ഇൻപുട്ട്: Port for connecting the EC sensor.

5. സജ്ജീകരണം

5.1. ബാറ്ററി ഇൻസ്റ്റാളേഷൻ

  1. മീറ്ററിന്റെ പിൻഭാഗത്ത് ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ കണ്ടെത്തുക.
  2. Use a coin or screwdriver to open the cover.
  3. ശരിയായ പോളാരിറ്റി (+/-) ഉറപ്പാക്കിക്കൊണ്ട് രണ്ട് (2) AA ബാറ്ററികൾ ഇടുക.
  4. വാട്ടർപ്രൂഫ് സീൽ നിലനിർത്താൻ ബാറ്ററി കമ്പാർട്ട്മെന്റ് കവർ സുരക്ഷിതമായി അടയ്ക്കുക.

The meter is designed for over 500 hours of continuous operation on a fresh set of batteries.

5.2. സെൻസർ കണക്ഷൻ

  1. Align the connector of the EC sensor (9383-10D) with the sensor input port on the top right side of the meter.
  2. Gently push and twist the connector clockwise until it is securely fastened. Do not overtighten.
  3. Ensure the sensor is clean and free from air bubbles before use.

6. പ്രവർത്തന നിർദ്ദേശങ്ങൾ

6.1. പവർ ഓൺ/ഓഫ്

അമർത്തുക ശക്തി button (bottom right) to turn the meter on or off.

6.2. കാലിബ്രേഷൻ

Regular calibration ensures accurate measurements. The EC220-K supports automatic and manual calibration for EC, TDS, and Salinity.

  1. അമർത്തുക CAL കാലിബ്രേഷൻ മോഡിൽ പ്രവേശിക്കാൻ ബട്ടൺ.
  2. Follow the on-screen prompts to select the appropriate calibration standard solution.
  3. Immerse the sensor in the standard solution. The meter will detect the solution and perform automatic calibration.
  4. For manual calibration or adjustment of reference temperature, temperature coefficient, or cell constant, use the സെറ്റ് ബട്ടണും അമ്പടയാള കീകളും.
  5. Confirm calibration with the ഇഎൻടി ബട്ടൺ.

6.3. അളവ്

After calibration, the meter is ready for measurement.

  1. അമർത്തുക MEAS button to enter measurement mode.
  2. അമർത്തുക മോഡ് button to cycle through measurement parameters: Conductivity (EC), Total Dissolved Solids (TDS), Salinity (SAL), and Resistivity (RES). Temperature is displayed concurrently.
  3. s-ലേക്ക് സെൻസർ മുക്കുകample solution. Ensure the sensor tip is fully submerged and free of air bubbles.
  4. Wait for the reading to stabilize. The 'Auto Stable' feature will indicate when the reading is stable.
  5. The meter can display temperature in both Celsius (°C) and Fahrenheit (°F), adjustable in the settings.

6.4. Data Logging and Memory

The LAQUA EC220-K can store up to 500 data points.

  • To save a measurement, press the ഇഎൻടി button when a stable reading is displayed.
  • വീണ്ടുംview സംഭരിച്ച ഡാറ്റ, അമർത്തുക ഡാറ്റ button. Use the arrow keys to navigate through the records.
  • The 'Auto Log Data' feature can be enabled in the settings for automatic data recording at set intervals.

6.5. ക്രമീകരണങ്ങൾ

Access advanced settings by pressing the സെറ്റ് button. Here you can adjust:

  • റഫറൻസ് താപനില
  • താപനില ഗുണകം
  • സെൽ കോൺസ്റ്റൻ്റ്
  • Measurement Units (e.g., µS/cm, mS/cm, ppm, ppt, °C, °F)
  • TDS Conversion Factor (four curves available)
  • Salinity Calibration Curves (two curves available)
  • ബാക്ക്ലൈറ്റ് ദൈർഘ്യം
  • Auto Power Off settings

7. പരിപാലനം

7.1. മീറ്റർ വൃത്തിയാക്കൽ

Wipe the meter body with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the battery compartment is sealed before cleaning.

7.2. Sensor Care and Cleaning

  • Rinse the sensor thoroughly with distilled or deionized water after each use.
  • If the sensor is visibly dirty or readings become erratic, clean it according to the sensor-specific instructions provided with the electrode.
  • Store the sensor in its protective cap with appropriate storage solution, if specified, to prevent drying out.

7.3. സംഭരണം

When not in use for extended periods, remove the batteries from the meter. Store the meter and sensor in the provided carrying case in a cool, dry place.

8. പ്രശ്‌നപരിഹാരം

പ്രശ്നംസാധ്യമായ കാരണംപരിഹാരം
മീറ്റർ ഓൺ ആകുന്നില്ല.മരിച്ചതോ തെറ്റായി ഇൻസ്റ്റാൾ ചെയ്തതോ ആയ ബാറ്ററികൾബാറ്ററികൾ മാറ്റി സ്ഥാപിക്കുക, ശരിയായ പോളാരിറ്റി ഉറപ്പാക്കുക.
കൃത്യമല്ലാത്ത വായനകൾSensor dirty or needs calibration; incorrect settingsClean sensor, recalibrate the meter, check reference temperature and cell constant settings.
Display shows 'Err' or '---'Sensor not connected; measurement out of rangeEnsure sensor is securely connected. Check if the sample is within the meter's measurement range.
മന്ദഗതിയിലുള്ള പ്രതികരണ സമയംSensor contamination or agingClean the sensor thoroughly. If the problem persists, the sensor may need replacement.

9 സ്പെസിഫിക്കേഷനുകൾ

പരാമീറ്റർവിശദാംശങ്ങൾ
മോഡൽലാക്വ ഇസി220-കെ
അളക്കൽ പാരാമീറ്ററുകൾConductivity (EC), Resistivity (RES), Salinity (SAL), Total Dissolved Solids (TDS), Temperature
വാട്ടർപ്രൂഫ്/ഡസ്റ്റ് പ്രൂഫ് റേറ്റിംഗ്IP67
ഡിസ്പ്ലേ തരംബാക്ക്‌ലൈറ്റുള്ള ഇഷ്ടാനുസൃത LCD
പവർ ഉറവിടം2 x AA ബാറ്ററികൾ
ബാറ്ററി ലൈഫ്500 മണിക്കൂറിലധികം
ഡാറ്റ മെമ്മറി500 ഡാറ്റ പോയിൻ്റുകൾ വരെ
കാലിബ്രേഷൻAutomatic and Manual EC Calibration, 4 TDS curves, 2 Salinity curves
ക്രമീകരിക്കാവുന്ന പരാമീറ്ററുകൾReference Temperature, Temperature Coefficient, Cell Constant, Measurement Units
ഹൗസിംഗ് മെറ്റീരിയൽPlastic (Shock and scratch-resistant non-slip casing)
ഉൾപ്പെടുത്തിയ ഘടകങ്ങൾമീറ്ററും സെൻസറും (9383-10D)
ഇനത്തിൻ്റെ ദൈർഘ്യം7 ഇഞ്ച്
നിർമ്മാതാവ്ഹോറിബ

10. വാറൻ്റിയും പിന്തുണയും

HORIBA LAQUA EC220-K മീറ്ററിൽ ഒരു 3 വർഷത്തെ വാറൻ്റി from the date of purchase, covering manufacturing defects. The sensor may have a separate warranty period; please refer to the sensor's specific documentation.

For technical support, service, or warranty claims, please contact HORIBA customer service or visit the official ആമസോണിലെ HORIBA സ്റ്റോർ അല്ലെങ്കിൽ അവരുടെ ഉദ്യോഗസ്ഥൻ webബന്ധപ്പെടാനുള്ള വിവരങ്ങൾക്കുള്ള സൈറ്റ്.

അനുബന്ധ രേഖകൾ - ലാക്വ ഇസി220-കെ

പ്രീview HORIBA LAQUA 200 സീരീസ് ഹാൻഡ്‌ഹെൽഡ് വാട്ടർ ക്വാളിറ്റി മീറ്ററുകൾ: സവിശേഷതകൾ, സ്പെസിഫിക്കേഷനുകൾ, ആക്സസറികൾ
HORIBA LAQUA 200 സീരീസ് ഹാൻഡ്‌ഹെൽഡ് വാട്ടർ ക്വാളിറ്റി മീറ്ററുകൾ കണ്ടെത്തൂ. IP67 വാട്ടർപ്രൂഫിംഗ്, വലിയ LCD ഡിസ്‌പ്ലേകൾ, കരുത്തുറ്റ നിർമ്മാണം എന്നിവയുൾപ്പെടെ pH, ORP, കണ്ടക്ടിവിറ്റി, ഡിസോൾവ്ഡ് ഓക്സിജൻ, TDS, ലവണാംശം, റെസിസ്റ്റിവിറ്റി മീറ്ററുകൾ എന്നിവയ്‌ക്കായുള്ള സ്പെസിഫിക്കേഷനുകൾ ഉൾപ്പെടെ PH210, PH220, EC210, EC220, DO210, DO220, PC210, PC220, PD210, PD220 എന്നീ മോഡലുകൾ ഈ ബ്രോഷറിൽ വിശദമായി പ്രതിപാദിക്കുന്നു. ലഭ്യമായ ആക്‌സസറികളും പരിഹാരങ്ങളും പര്യവേക്ഷണം ചെയ്യുക.
പ്രീview HORIBA LAQUA-EC210/EC220 കണ്ടക്ടിവിറ്റി മീറ്റർ ഇൻസ്ട്രക്ഷൻ മാനുവൽ
HORIBA LAQUA-EC210, LAQUA-EC220 ഹാൻഡ്‌ഹെൽഡ് കണ്ടക്ടിവിറ്റി മീറ്ററുകൾക്കായുള്ള സമഗ്രമായ നിർദ്ദേശ മാനുവൽ. കൃത്യമായ ജല ഗുണനിലവാര വിശകലനത്തിനായി പ്രവർത്തനം, കാലിബ്രേഷൻ, സജ്ജീകരണം, പരിപാലനം, ട്രബിൾഷൂട്ടിംഗ് എന്നിവ ഉൾക്കൊള്ളുന്നു.
പ്രീview ഹൊറിബ ലാക്വ സീരീസ് വാട്ടർ ക്വാളിറ്റി മീറ്റർ കാറ്റലോഗ് 2025-2026
HORIBA LAQUA സീരീസ് pH, ചാലകത, ORP, DO മീറ്ററുകളുടെ സമഗ്രമായ കാറ്റലോഗ്. വൈവിധ്യമാർന്ന ജല ഗുണനിലവാര അളക്കൽ ആപ്ലിക്കേഷനുകൾക്കായി വിവിധ ഇലക്ട്രോഡ് ഓപ്ഷനുകളുള്ള ബെഞ്ച്ടോപ്പ്, പോർട്ടബിൾ, കോംപാക്റ്റ് മോഡലുകൾ എന്നിവ ഇതിൽ ഉൾപ്പെടുന്നു.
പ്രീview HORIBA LAQUA PD210/PD220 നിർദ്ദേശ മാനുവൽ: pH, ORP, അലിഞ്ഞുചേർന്ന ഓക്സിജൻ മീറ്റർ
ഈ മാനുവൽ HORIBA LAQUA 200 സീരീസ് ഹാൻഡ്‌ഹെൽഡ് വാട്ടർ ക്വാളിറ്റി മീറ്ററുകളുടെ പ്രവർത്തനം വിവരിക്കുന്നു, മോഡലുകൾ LAQUA-PD210, LAQUA-PD220. ഇത് ഉൽപ്പന്നത്തെ മുഴുവൻ ഉൾക്കൊള്ളുന്നുview, pH, ORP, അലിഞ്ഞുചേർന്ന ഓക്സിജൻ അളവുകൾക്കായുള്ള അടിസ്ഥാന പ്രവർത്തനങ്ങൾ, കാലിബ്രേഷൻ, ഡാറ്റ മാനേജ്മെന്റ്, സജ്ജീകരണം, പരിപാലനം, ട്രബിൾഷൂട്ടിംഗ്.
പ്രീview HORIBA LAQUA 2000 സീരീസ് ബെഞ്ച്‌ടോപ്പ് വാട്ടർ ക്വാളിറ്റി മീറ്ററുകൾ: സവിശേഷതകൾ, സ്പെസിഫിക്കേഷനുകൾ & ആക്‌സസറികൾ
PH2000, EC2000, ION2000, DO2000, PC2000, PD2000 മോഡലുകൾ ഉൾപ്പെടെയുള്ള HORIBA LAQUA 2000 സീരീസ് ബെഞ്ച്‌ടോപ്പ് വാട്ടർ ക്വാളിറ്റി മീറ്ററുകളുടെ സമഗ്ര ശ്രേണി കണ്ടെത്തൂ. ലബോറട്ടറിയിലും ഫീൽഡ് ആപ്ലിക്കേഷനുകളിലും കൃത്യമായ ജല വിശകലനത്തിനായി അവയുടെ നൂതന സവിശേഷതകൾ, കൃത്യമായ സ്പെസിഫിക്കേഷനുകൾ, ലഭ്യമായ ആക്‌സസറികൾ എന്നിവ ഈ ബ്രോഷറിൽ വിശദമായി പ്രതിപാദിക്കുന്നു.
പ്രീview HORIBA LAQUA 2000 സീരീസ് ബെഞ്ച്‌ടോപ്പ് മീറ്റർ ഇൻസ്ട്രക്ഷൻ മാനുവൽ
HORIBA LAQUA 2000 സീരീസ് ബെഞ്ച്‌ടോപ്പ് pH/ORP/ION/ചാലകത മീറ്ററുകൾക്കായുള്ള (LAQUA-PH2000, LAQUA-ION2000, LAQUA-EC2000, LAQUA-PC2000) സമഗ്രമായ നിർദ്ദേശ മാനുവൽ. പ്രവർത്തനം, കാലിബ്രേഷൻ, സജ്ജീകരണം, പരിപാലനം, ട്രബിൾഷൂട്ടിംഗ് എന്നിവ ഉൾക്കൊള്ളുന്നു.