1. ആമുഖം
The Extech DO600 Waterproof ExStik II Dissolved Oxygen Meter is designed for precise measurement of dissolved oxygen levels. It displays oxygen concentration as a percentage of saturation (0 to 200%) or as a concentration in parts per million (ppm) or milligrams per liter (mg/L) (0 to 20.00 ppm/mg/L). Key features include adjustable altitude compensation, adjustable salinity compensation, and an analog bargraph for trend indication. This manual provides instructions for the proper setup, operation, and maintenance of your DO600 meter.
2. പാക്കേജ് ഉള്ളടക്കം
അൺപാക്ക് ചെയ്യുമ്പോൾ എല്ലാ ഇനങ്ങളും നിലവിലുണ്ടെന്നും നല്ല നിലയിലാണെന്നും ഉറപ്പാക്കുക:
- Extech DO600 Dissolved Oxygen Meter
- ഡിഒ ഇലക്ട്രോഡ്
- Protective Sensor Cap
- Spare Membrane Cap
- Electrolyte Solution
- Four 3V CR-2032 Batteries (pre-installed or included separately)
- 48-inch (1.2m) Neck Strap

Figure 1: Extech DO600 meter with its accessories, including the electrode, caps, electrolyte, and neck strap.
3. Meter Components

ചിത്രം 2: മുൻഭാഗം view of the Extech DO600 meter.
- LCD ഡിസ്പ്ലേ: Shows dissolved oxygen readings, temperature, and other indicators.
- മോഡ്/ഹോൾഡ് ബട്ടൺ: Toggles between measurement modes and holds the current reading.
- CAL/RECALL Button: Initiates calibration and recalls stored data.
- ഓൺ/ഓഫ് ബട്ടൺ: മീറ്റർ ഓൺ അല്ലെങ്കിൽ ഓഫ് ആക്കുന്നു.
- ഇലക്ട്രോഡ്: Sensor for dissolved oxygen measurement.
- Protective Sensor Cap: Covers and protects the electrode.
4. സജ്ജീകരണം
4.1. ബാറ്ററി ഇൻസ്റ്റാളേഷൻ
The DO600 meter uses four 3V CR-2032 batteries. These are typically pre-installed. If replacement is necessary:
- Twist off the battery compartment cap located at the top of the meter.
- Remove the old batteries, noting their polarity.
- Insert new CR-2032 batteries with the correct polarity.
- വാട്ടർപ്രൂഫ് സീൽ നിലനിർത്താൻ ബാറ്ററി കമ്പാർട്ട്മെന്റ് ക്യാപ്പ് സുരക്ഷിതമായി മാറ്റിസ്ഥാപിക്കുക.
4.2. Electrode Preparation
Before first use or after prolonged storage, the DO electrode requires preparation:
- Unscrew the protective sensor cap from the electrode.
- Fill the spare membrane cap with the provided electrolyte solution. Ensure no air bubbles are trapped.
- Carefully screw the membrane cap onto the electrode, ensuring the membrane is taut and free of wrinkles.
- Allow the electrode to stabilize for approximately 15-30 minutes in the electrolyte solution before calibration.
5. പ്രവർത്തന നിർദ്ദേശങ്ങൾ
5.1. പവർ ഓൺ/ഓഫ്
അമർത്തുക ഓൺ/ഓഫ് button to power the meter on or off.
5.2. കാലിബ്രേഷൻ
Regular calibration ensures accurate readings. Calibrate the meter in air-saturated water or a known standard solution.
- Ensure the electrode is prepared as described in Section 4.2.
- Immerse the electrode in air-saturated water (e.g., tap water stirred vigorously for several minutes) or a DO standard solution.
- അമർത്തിപ്പിടിക്കുക CAL/RECALL button until "CAL" appears on the display.
- The meter will automatically recognize the calibration point and adjust. Wait for the reading to stabilize and for the calibration indicator to confirm completion.
5.3. അളവുകൾ എടുക്കൽ
To measure dissolved oxygen:
- മീറ്ററിൽ പവർ.
- ഇലക്ട്രോഡ് s-ൽ മുക്കുക.ample solution. Ensure the membrane is fully submerged and free of air bubbles.
- Gently stir the electrode in the sample to ensure proper flow across the membrane.
- Wait for the reading on the LCD display to stabilize. The display will show both dissolved oxygen and temperature.
- അമർത്തുക മോഡ്/ഹോൾഡ് button to cycle between % saturation and ppm (mg/L) readings.
- To hold a reading, briefly press the മോഡ്/ഹോൾഡ് ബട്ടൺ. റിലീസ് ചെയ്യാൻ വീണ്ടും അമർത്തുക.

Figure 3: The Extech DO600 meter being used to measure dissolved oxygen in a laboratory.
5.4. Altitude Compensation
Dissolved oxygen saturation levels are affected by altitude. The DO600 allows for manual adjustment:
- Refer to the meter's full instruction manual for specific steps to enter and adjust altitude compensation (0 to 20,000 feet in 1,000 ft. increments).
5.5. Salinity Compensation
Salinity also influences dissolved oxygen levels. The DO600 provides adjustable salinity compensation:
- Consult the meter's comprehensive manual for detailed instructions on setting salinity compensation (0 to 50 ppt).
6. പരിപാലനം
6.1. Electrode Cleaning and Storage
To maintain electrode performance:
- ഓരോ ഉപയോഗത്തിനും ശേഷം വാറ്റിയെടുത്ത വെള്ളം ഉപയോഗിച്ച് ഇലക്ട്രോഡ് കഴുകുക.
- If the membrane becomes fouled, gently clean it with a soft cloth or cotton swab. Replace the membrane cap if it is damaged or heavily contaminated.
- When storing, ensure the protective sensor cap is filled with electrolyte solution to keep the membrane hydrated.
6.2. ബാറ്ററി മാറ്റിസ്ഥാപിക്കൽ
ഡിസ്പ്ലേയിൽ ബാറ്ററി കുറവാണെന്ന് കാണിക്കുമ്പോൾ ബാറ്ററികൾ മാറ്റിസ്ഥാപിക്കുക. ബാറ്ററി ഇൻസ്റ്റാളേഷൻ നിർദ്ദേശങ്ങൾക്ക് സെക്ഷൻ 4.1 കാണുക.
7. പ്രശ്നപരിഹാരം
| പ്രശ്നം | സാധ്യമായ കാരണം | പരിഹാരം |
|---|---|---|
| മീറ്റർ ഓൺ ആകുന്നില്ല. | ഡെഡ് ബാറ്ററികൾ; തെറ്റായ ബാറ്ററി ഇൻസ്റ്റാളേഷൻ. | ബാറ്ററികൾ മാറ്റിസ്ഥാപിക്കുക; ബാറ്ററി പോളാരിറ്റി പരിശോധിക്കുക |
| Inaccurate or unstable readings | Electrode not calibrated; Fouled membrane; Air bubbles on membrane; Insufficient electrolyte | Perform calibration; Clean/replace membrane; Remove air bubbles; Refill electrolyte |
| വായനയൊന്നും പ്രദർശിപ്പിച്ചിട്ടില്ല | Electrode not connected properly; Damaged electrode | Ensure electrode is securely attached; Replace electrode if damaged |
8 സ്പെസിഫിക്കേഷനുകൾ
| പരാമീറ്റർ | മൂല്യം |
|---|---|
| മോഡൽ | DO600 |
| Dissolved Oxygen Range (Concentration) | 0 മുതൽ 20.00 ppm (mg/L) |
| Dissolved Oxygen Range (Saturation) | 0 മുതൽ 200.0% വരെ |
| താപനില പരിധി | Not explicitly stated for display, but upper rating is 50 ℃ |
| ഉയരം നഷ്ടപരിഹാരം | 0 to 20,000 feet (in 1,000 ft. increments) |
| ലവണാംശ നഷ്ടപരിഹാരം | 0 മുതൽ 50 വരെ |
| പവർ ഉറവിടം | Four 3V CR-2032 batteries |
| അളവുകൾ | 6.9 x 1.6 x 1.4 ഇഞ്ച് (17.5 x 4.1 x 3.6 സെ.മീ) |
| ഇനത്തിൻ്റെ ഭാരം | 3.87 ഔൺസ് (0.11 കി.ഗ്രാം) |
| വാട്ടർപ്രൂഫ് റേറ്റിംഗ് | Yes (implied by "Waterproof ExStik II") |

Figure 4: Physical dimensions of the Extech DO600 meter. Note: The image shows 7.7 inches, while product specifications state 6.9 inches. Refer to the latest product documentation for precise measurements.
9. ഉൽപ്പന്ന വീഡിയോകൾ
നൽകിയിരിക്കുന്ന ഡാറ്റയിൽ ഈ ഇനത്തിനായി വിൽപ്പനക്കാരനിൽ നിന്നുള്ള ഔദ്യോഗിക ഉൽപ്പന്ന വീഡിയോകളൊന്നും കണ്ടെത്തിയില്ല.
10. വാറൻ്റിയും പിന്തുണയും
For warranty information, technical support, or to purchase accessories and replacement parts, please refer to Extech's official webസൈറ്റിൽ ബന്ധപ്പെടുക അല്ലെങ്കിൽ അവരുടെ ഉപഭോക്തൃ സേവന വിഭാഗവുമായി ബന്ധപ്പെടുക. കൂടുതൽ വിവരങ്ങൾ നിങ്ങൾക്ക് ഇവിടെ കണ്ടെത്താനാകും ആമസോണിലെ എക്സ്ടെക് സ്റ്റോർ.